11.1 Technical specifications

Analyzer SSM6000
Gas type
Measuring range
0 ... 100
Vol.%.
CH
4
0 ... 5.000
ppm
H
S
2
0 ... 1.000
ppm
0 ... 1000
ppm
H
2
0 ... 25
Vol.%.
O
2
0 ... 100
Vol.%.
CO
2
Display and measured-value displays
Analog outputs
Digital outputs
Digital inputs
Interfaces
Flowmeter / control valve
Measuring-gas pump / valves
Housing/dimensions
System housing
Housing dimensions (WxHxD) - without
Build-in dimensions (WxHxD)
Weight
Protection
Housing color / door frame
Forced ventilation / housing rinsing
Requirements for the place of installation
Operating temperature
Lagertemperatur
Rel. humidity
Air pressure
Measuring input and output conditions, calibration gases
Input dew point of the gas to be measured
Measuring-gas temperature
Pressure at the measuring-gas input
Measuring-gas output pressure
Test gas mixture for calibration (example)
800C-07.07.EN
Resolution
Precision
0,1
Vol.%.
±2% FS
1 / 5
ppm
±5% FS
1
ppm
±5% FS
1
ppm
±5% FS
0.1
Vol.%.
±2% FS
0.1
Vol.%.
±2% FS
4-digit LED measured-value display, status displays for each channel
4-line LC display, illuminated
4–20 mA (1 output oper measuring component, linearized)
aktiv, common reference potential, max. burden 550Ω)
Status signals: stand-by / measurement active / setup mode
Limit-value alarms: violation of upper limit 1x per measuring component + violation of lower limit
for CH
and O
4
2
Valve outputs: measuring-point switch / test gas valve
transistor, open collector type, external supply 35VDC / 750mA max.
external start-up of measureing / choice of the measuring point
optocoupler, typ. switching threshold 10VDC, external supply 24VDC max.
RS 232, digital output of measured values and program updates
Profibus / CAN bus (option)
Rotmater 7-70 l/h, gas connections Viton / needle valve , PA
once installed, quantity refers to specification
Steel sheet housing with front door and inspection window
300 x 400 x 185 mm
500 x 700 x 500 mm
Approx. 11-15 kg (abhängig von der Geräteausführung)
IP 20
RAL 7035
Housing fan 12VDC / 56 m³/h free-blowing - only if option measuring cooler
+10 to +40°C
-25 to +50°C
<=75% annual mean; minor and seldom condensation in de-energized condition
850 ... 1100 hPa
max. 40°C (with measuring cooler)
min. 5°C below ambient temperature limit (without measuring cooler)
80°C max.
-200 ... +200 hPa
Exhaust air must be discharged into the outside atmosphere in a non-pressurized condition /
discharge hoses should be as short as possible
40Vol.%. of CO
/ 300ppm H
2
SSM 6000C User Manual
Other
Measuring method
Temperature and pressure
Two-beam IR
Electrochemical
Dilution stages 1:200/40/10/0 (MP4)
Electrochemical
Without Dilution (MP5)
Electrochemical
Without Dilution (MP5) ,Instead of CO
Electrochemical
Temperature and pressure
Two-beam IR
S / 0 Vol.%. oif O
/ 800ppm H
2
2
2
TECHNICAL SPECIFICATIONS
/ rest CH
4
2
61