Installing The Gas Analyzer; Unpacking And Assembling The Gas Analyzer; Electrical Connections - PRONOVA SSM 6000 User Manual

Hide thumbs Also See for SSM 6000:
Table of Contents

Advertisement

SSM 6000
CONT
3.5

Installing the gas analyzer

3.5.1

Unpacking and assembling the gas analyzer

Unpacking the gas
Step
analyzer
1
2
3
The analyzer was carefully packed for shipment. Before installing and setting into operation, check for possible transport damage. In
the case of transport damage with signs of incorrect handling, have the damage recorded by the carrier (rail, post, mover) within
seven days.
Make sure not to lose the accessories enclosed (refer to the "Scope of delivery" section).
The transport box and the rubber foam parts should be kept for future shipment of the analyzer (calibration and analyzer inspection).
The analyzer should be mounted on a vertical, stable wall using sufficiently sized fastening elements. When selecting the place of
installation, please remember that free space of at least 25 cm is required underneath the device for electrical connections and
hoses. A distance of at least 15 cm should be maintained between the right hand side of the device and the nearest wall in order to
ensure sufficient air circulation and to enable the replacement of the housing filter.
3.5.2

Electrical connections

All electrical connections are located on the underside of the analyzer.
Fig.: Analyzer view from below (standard)
800C-07.07.EN
Action
Carefully remove the transport packaging of the analyzer, and store the analyzer in a clean place.
Remove the rubber foam parts from the analyzer system.
Visually check the analyzer system for transport damage before assembling.
SSM 6000C User Manual
HARDWARE INSTALLATION
25