SSM 6000
CONT
Proceed in accordance with the applicable national safety regulations for the construction and operation of
electrical systems as well as the safety information below!
The connection between a protective earth connection and a protective earth conductor must be made before any
other connections are made!
The gas analyzer can become a hazard if the protective earth conductor is interrupted inside or outside the gas
analyzer or if the protective conductor connection is disconnected!
Live parts can be exposed when covers are opened or when parts are removed unless this can be carried out
without the use of tools. Furthermore, connection points can also be live.
Voltage supply
The device comes with a power cable with earthed-contact plug and may only be connected to an earthed-contact receptacle. The
protective effect may not be rendered ineffective by using an extension cable without PE conductor.
We recommend installing a mains disconnector switch or a switched receptacle in the power supply circuit near the gas analyzer in
order to be able to disconnect all poles of the gas analyzer from the power supply, if necessary. The mains disconnector switch must
be marked in a manner that clearly shows the equipment to be disconnected.
Mains voltage:
Power consumption:
A QUICKON type cable screw connection is foreseen for the power supply for the analyzer. In the case of installation without an
earthed-contact plug, the analyzer is connected directly using the screw connection according to the following installation
instructions.
The screw-type cable connection is suitable for the following power cables:
Cable specifications:
Connecting the voltage supply within the device by the customer is not permitted by the manufacturer and will render
the guarantee invalid.
800C-07.07.EN
230 VAC / 50 Hz (option 115VAC / 60 Hz), quod vide appliance rating plate
85 VA max.
Light to medium conduits, such as. H03VV-F / H05VV-F, number of poles = 3 (L/N/PE)
Flexible cross-section 0.75 to 1.5 mm² / conductor configuration according to VDE 0295 = class 2 to 5
Wire diameter, including insulation, up to 3mm / outside diameter 5.6 to 9 mm
SSM 6000C User Manual
HARDWARE INSTALLATION
26