Sensor Data - PRONOVA SSM 6000 User Manual

Hide thumbs Also See for SSM 6000:
Table of Contents

Advertisement

SSM 6000
CONT
5.3.3

Sensor data

SENSOR DATA:
_ ACTUAL
_ HISTORY
Selecting the Sensor data
the measuring-gas input starts flowing through the CH
from 0 to 1023 dig.
The theoretical display can hence be calculated as follows at a defined calibration gas concentration.
Gas
[ ]
=
Display
dig
Measuring
The above equation does not have to be fulfilled exactly. The equation neglects, for example, the fact that the signal should feature a
slight offset with zero gas (living zero point). The actual sensor signals should be within the limits shown below.
Actual sensor signals [dig]
SENSOR:
30.0 C
1: 600 4: 400 7:
0
2: 400 5: 434 8:
3: 780 6:1010 9:
In order to determine the zero point of the sensor, the biogas hose must be disconnected from the detonation protection
1)
unit, so that ambient air is drawn in through the measuring-gas input.
A calibration gas / biogas with the following composition was used as a basis for the sensor signals indicated:
2)
59.9 percent by volume of CH
By selecting the Sensor data
History – sensor signals of the last measurements [dig]
LAST MEASURES:
_20.01. 12:00
_21.01. 12:00
_22.01. 12:00
MEASURE 01.05. 17:57
A:
10 D: 400 G:
30
B: 600 E: 040 H:
8
C:
52 F: 780 I: 400
800C-07.07.EN
The Sensor data menu serves the functional checking of the sensors and thereby supports remote
diagnosis by the manufacturer.
The current sensor signals and the signals of the last three measurements (HISTORY) can be
displayed.
Actual activates the measuring-gas pumps in the analyzer whereupon the gas currently available at
gas system. The display shows the current sensor signals in number blocks
4
concentrat
ion
1023
range
1
Sensor signal CH
2
Sensor signal H
3
Sensor signal O
4
Sensor signal CO
Sensor signal H
5
Cooler temperature
6
Pressure sensor [mbar]
/ 300 ppm H
S,/ 850 ppm H
4,
2
History menu item, you can check the sensor signals of the last three measurements.
A
Sensor signal CH
B
C
Sensor signal H
D
E
Dilution stage H
F
Sensor signal O
G
H
Sensor signal CO
I
H
Sensor signal H
I
SSM 6000C User Manual
The measuring range totals 100 percent by volume for CH
dig
and CO
. The display of the sensor signal for H
2
on the dilution stage. With the 1:40 dilution stage, the
measuring range totals 1000 ppm.
Ambient air
0001 - 0010
4
S
0001 - 0075
2
0650 - 1020
2
0001 - 0010
2
(instead of CO
)
0001 - 0060
2
2
0950 - 1050
/ 0 percent by volume of O
2
Zero gas
4
Biogas / calibration gas
S
Zero gas
2
Biogas / calibration gas
S
Biogas / calibration gas
2
Test gas
2
Zero gas
Zero gas
2
Biogas / calibration gas
(instead of CO
) Zero gas
2
2
Biogas / calibration gas
MENU AND FUNCTION OVERVIEW
S depends
2
Calibration gas
1)
2)
0525 – 0675
0001- 0075
0001 – 0050
0325 – 0475
0750 – 0900
390
0950 – 1050
/ 40 percent by volume of CO
2
Sensor signals
0001 - 0020
0525 - 0675
0001 - 0075
< 1023
0, 10, 40 or 200
0001 - 0050
0650 - 1023
0001 - 0020
0325 - 0475
0001 - 0060
0750 - 0900
4
.
2
2)
44