PRONOVA SSM 6000 User Manual page 39

Hide thumbs Also See for SSM 6000:
Table of Contents

Advertisement

SSM 6000
CONT
Output on RS232
Via the serial interface, the measured values, including a time stamp, can be transmitted to a PC or to the general system controller
and evaluated using a suitable program. In this way, trends and dependencies between the individual measured values can be
identified.
The values of the last measuring process, including date and time information, are continuously output via the interface, depending
on the
selected, or can be exported as a block at a later point in time from the data logger of the analyzer. The data
RS232 interval
logger of the SSM 6000 can store around 4096 measurements / data records.
DATALOGGER:
2354
_ OUTPUT TO RS232
_ OUTPUT TO LCD
_ RS232 INTERVALL
The measuring data is output in the csv/txt format via the RS232 interface as follows:
Date Time / CH
/ H
4
16.07.03 11:50;60,6;0282;00,0;40,4;23,4;1013;1;
16.07.03 11:59;60,6;0284;00,0;40,4;23,6;1013;2;
16.07.03 12:52;00,0;0019;00,0;00,0;23,7;1014;1;
16.07.03 13:02;00,0;0001;20,8;00,0;23,8;1014;4;
As a precondition for the transmission of data from the SSM 6000 to the PC, the terminal program must first be configured. This
procedure is described here using the "HyperTerminal" program under MS Windows as an example. However, the properties of the
serial interface can also be transferred to other terminal programs.
Configure HyperTerminal
The terminal program is accessed via the start bar
HyperTerminal icon whereupon you can now set up a new connection.
800C-07.07.EN
RS232 INTERVALL:
_ EVERY
0001 s
S / O
/ CO
/ Device temperature / Air pressure / Measuring-point No.
2
2
2
Programs
SSM 6000C User Manual
Accessories
Communication
In this dialog field, enter a name (such as
connection and select any icon you like.
Thereafter, click
to confirm.
OK
MENU AND FUNCTION OVERVIEW
. Click the
HyperTerminal
) for the new
SSM 6000