SSM 6000
CONT
Transmitting measured data
As a precondition for transmitting measured values from the SSM 6000, the analyzer must be connected via the RS232 interface to
the serial interface (COM1 or COM2) of the PC.
Note:
Use a suitable serial cable (1:1 connector layout) for this connection. The exact connector layout of the interface is shown
in the terminal diagrams in the appendix. Do not use a null modem cable because these cables feature an incorrect
connector layout, so that data transmission is not possible.
At first, open the previously configured HyperTerminal connection from the start menu:
SSM 6000 or via the link on the desktop.
HyperTerminal
The HyperTerminal window SSM 6000 is opened:
In order to avoid a major loss of data in the case of a defect of the device, we recommend exporting the data at regular
intervals. The measured data is also exported and archived by the manufacturer each time service work is carried out.
This data is then available on demand.
In order to export the stored measured data from the data logger, select
and press <enter> in order to start the data transmission process.
OUTPUT TO RS232
The progress of the transmission process can be monitored on the screen of the PC and on the LC display.
DATALOGGER:
2345
_ OUTPUT TO RS232
_ OUTPUT TO LCD
_ RS232 INTERVALL
800C-07.07.EN
HyperTerminal immediately and automatically establishes the connection
to the SSM 6000.
Select the "Transmission – Record text" menu item.
During the next step, enter any file name, such as measurement_050322,
and select the target directory in which the measured-value file is to be
stored.
Click
in order to confirm the entry.
Start
DEVICE OPTIONS – DATA OUTPUT – DATA LOGGER –
OUTPUT TO RS232:
_ START UPLOAD
In order to terminate the transmission process and save the data, select
the
Transmission – Record text – Exit
The stored data can now be imported to a suitable program and evaluated
there.
We recommend selecting a new file name for the transmission of new
measuring data.
SSM 6000C User Manual
MENU AND FUNCTION OVERVIEW
Programs
Accessories
menu item.
Communication
41