SSM 6000
CONT
6.2.3

Setting the measuring interval

For the purpose of discontinuous H
set into operation. In the case of the other gas types to be measured, the only important point is to edit an interval other than zero
because these gases are measured continuously.
SETUP MODE:
_ MEASUREMENT
_ CALIBRIATION
_ SYSTEM
Not every interval is permitted. The interval must be a divisor of 1440, i.e. the number of minutes of a day. If the value
entered does not fulfill this requirement, the SSM 6000 automatically uses the nearest setting. With the setting selected
above, H
S is measured every 480 minutes = three times a day at 4:00 a.m., 12:00 noon, and 8:00 p.m.
2
However, the measuring interval should not be set at a value of less than 120 minutes. This corresponds to 12 measurements a day.
The reference time, "BZ", determines the time when the first H
measuring process is then started xxx minutes later, depending on the set H
The "Offset" selection (_OFS: 0000) enables the automatic starting of measuring processes in the intervals defined for the day with
the "minutes" entry edited to this effect.
If the device is controlled via the digital inputs or via the Profibus interface, an H
order to avoid conflicts with the automatic measuring interval.
6.2.4

Other settings

If required, further device configuration settings can be made as required. For a more detailed description, please refer to the "Menu
and function overview" section.
- Select the menu language, i.e. German / English
- Configuration of analog outputs 0 – 20 mA / 4 – 20 mA
- Setting the analog operating modes
- Measuring range of the analog output for H
- Setting the RS232 interval
800C-07.07.EN
S measurement, the measuring interval and the reference time must be set when the analyzer is
2
MEAS. OPTIONS:
_ MEAS. INTERVAL
_ ALERT VALUES
_ SINGLE MEAS.
S measuring process starts. After this measuring process, the next
2
S
2
SSM 6000C User Manual
MEAS.INTERVAL: [min]
_CH4:
240 _O2 : 240
_H2S:
480 _CO2: 240
_RT :12:00 _OFS:
S measuring interval.
2
S measuring interval of "9999" should be set in
2
Default setting: German
Default setting: 4 – 20 mA
Default setting: warm-up phase = XmA / rest = VAL
Default setting: 0 – 5000 / 0 – 1000 ppm H
Default setting: 1 second
SETTING INTO OPERATION
30
S
2
47