Setting Into Operation; Switching The Gas Analyzer On; Setting Operating Parameters; Setting Date And Time - PRONOVA SSM 6000 User Manual

Hide thumbs Also See for SSM 6000:
Table of Contents

Advertisement

SSM 6000
CONT

6 Setting into operation

When all the activities described in the "Installation" section have been completed and after all electrical connections and hose
connections have been once again checked, the analyzer can be set into operation.
6.1

Switching the gas analyzer on

In order to set the SSM 6000 into operation, proceed as follows:
Connect voltage supply.
-
Press the power button on the front panel in order to switch the analyzer on.
-
After power-on, the display of the analyzer shows the version number for around 5 seconds. This is followed by a warm-up phase of
around 30 minutes which ends automatically on expiration of this time.
Since the only initial operations carried out on the device are settings, the warm-up phase can be terminated by pressing the esc
key. The device is now in stand-by mode and parameterization can start in setup mode after operation of the key switch.
6.2

Setting operating parameters

In order to adjust the analyzer to the requirements of the biogas plant, certain settings must first be made with the analyzer in setup
mode. This mode is activated by operating the key switch on the control panel of the analyzer.
6.2.1

Setting date and time

First check and, if necessary, adjust the date and time. This setting is important for future activities which must be regularly triggered
at a certain point in time.
SETUP MODE:
_ MEASUREMENT
_ CALIBRIATION
_ SYSTEM
6.2.2

Setting alarm values

In this menu item, you can edit the alarm thresholds for the individual gas types which are specified in the operating instructions
issued by the plant manufacturer.
MEAS. OPTIONS:
_ MEAS. INTERVAL
_ ALERT VALUES
_ SINGLE MEAS
The values for CH
, O
and CO
4
2
The ex-works settings are non-binding recommendations. The manufacturer does not assume any liability for the alarm
values set! Please contact the supplier of your plant in order to identify the values to be set in order to avoid damage to
the equipment!
800C-07.07.EN
SYSTEM OPTIONS:
_ DATA OUTPUT
_ OPERATION MODE
_ DATE/TIME
ALERT VALUES:
_MEAS. GASES
_TEMPERATURES
are expressed in percent by volume, whilst the H
2
SSM 6000C User Manual
SETTING INTO OPERATION
DATE/TIME:
_ DATE
: 11:40:02
_ TIME
: 07.11.04
ALARM VALUES:
_CH4<
50
_O2 < 0,5
_CH4>
70
_O2 >
5
_H2S> 250
_CO2>
55
S value is expressed in ppm/vpm.
2
46