Replacing The Baffle Box - Weinmann homecare SOMNOcomfort 2 Servicing And Repair Instructions

Table of Contents

Advertisement

3. Remove the decoupling tube including the
insert.
4. Loosen the screw
72
5. Remove the display
Attaching the display
1. Glue new foam
to the display
61
2. Insert the new display.
3. Tighten the screw
72
4. Attach the decoupling tube.
– Connect the decoupling tube to the box.
– Insert the decoupling tube including the
insert.
– Tighten the screw
If the screws are too tight it may
Note:
become difficult to install the cover
5. Insert the main circuit board (chapter 7.6,
page 25, steps
to
1.
6. Close the unit (chapter 7.5, page 24).
7. Check that everything is working properly
(chapter 5., page 15).

7.10 Replacing the baffle box

1. Open the unit (chapter 7.4, page 23).
2. Remove the main circuit board ("7.6
Replacing the main circuit board" on
page 25, steps
to
1.
Remove the baffle box
1. Remove the decoupling tube
2. Remove the box
in an upwards motion.
53
28
Repair information and repair instructions
.
.
51
.
51
using the spring
.
79
but not too tight.
71 firmly,
.
5
).
4.
).
5.
from the box.
53
50
72
79
51
53
50

Hide quick links:

Advertisement

Table of Contents
loading