5.2.3 Checking the power supply

1. Connect the power supply
power source.
– If the LED lights up, the power supply is in
order.
2. Replace the power supply

5.2.4 Checking the housing

Check the housing's general condition.
If the housing is damaged in any way, replace
the respective part of the housing (see also
Chapter 7.12, Page 31 and Chapter 7.13,
Page 31).
5.2.5 Checking the control panel and indicators
1. Make sure that front panel foil of the control
panel is attached flush and tight. If this is not
the case, replace the foil (see chapter "7.7"
on page 26).
2.
You should now establish the power supply
by connecting the power supply
power supply cable
available power outlet.
3. Close the opening of the mask, e.g. using your
thumb or hand and hold it closed.
4. Push the on/off switch
on.
On the display, the total hours of operation
will be indicated for 3 seconds before the
current measured pressure will appear.
5. Press the Softstart button
change.
6. Switch the unit off using the
The display will show the date and the day's
length of therapy for a short while.
Subsequently only "0" and "A" or "M" will be
shown.