7.12 Replacing the housing, lower part

1. Open the unit (chapter 7.4, page 23).
2. Loosen the lock and pull the connection plug
from the lower part of the housing
64
upward motion.
3. Place the connecting line
pointing
downwards
housing
. The plug will lock in.
48
4. Close the unit (chapter 7.5, page 24).
5. Check that everything is working properly
(chapter 5., page 15).

7.13 Replacing the housing, upper part

1. Open the unit (chapter 7.4, page 23).
2. Remove the main circuit board (chapter 7.6,
page 25, steps
to
1.
3. Remove the display (chapter 7.9, page 27,
steps
to
).
1.
5.
4. Disassemble the adapter pressure gland
5. Remove the box
(chapter 7.10, page 28,
53
steps
to
).
1.
2.
You have now removed all of the parts. Now
begin to reassemble them back into the new upper
part of the housing
.
47
6. Attach front panel foil
Make sure the flat cable will be inserted into the
according opening of the housing cleanly without
any kinks.
7. Remove the protective foil from the new front
panel foil.
8. Insert the box
(chapter 7.10, page 28,
53
steps
to
).
1.
4.
9. Attach the adapter pressure gland
The nub of the silicone part must be inserted into
the appropriate cut-out of the plastic part (see
illustration).
10. Insert the display (chapter 7.9, page 27,
steps
to
).
2.
4.
11. Insert the main circuit board (chapter 7.6,
page 25, steps
to
1.
12. Close the unit (chapter 7.5, page 24).
in an
48
with the plug
64
into the socket of the new
).
5.
.
78
to the housing.
52
.
78
).
4.
48
64
78
Repair information and repair instructions
31