Labconco SteamScrubber 4010010 Series User Manual page 98

Glassware washers
Table of Contents

Advertisement

Step Skipped
Temp Sensor
Open buttons or
display on control
panel inoperable or
controls
malfunction
Detergent cup in
door fails to open
Poor drying
Segment Skipped by
SS
operator.
Temperature Sensor
TS
failure.
Microprocessor
memory error.
Keypad or display
defective.
Glassware
interfering with cup
door.
Malfunction of cup.
D-cup latch fouled.
Dry temperature too
low.
Flask, beaker, or
labware with
concave bottoms not
loaded correctly.
Recondensation.
Dry time too short.
Heater(s)
disconnected or
malfunctioned
Blower disconnected
or malfunctioned.
No action
required.
Check Sensor
Screen for temp
reading.
Detergent Cup
Test: with door
closed
Manually test door
operation.
Heater/s Test:
Dry Bower Test.
98
Program will drain
and begin next
programed
segment.
If no reading and
wiring connected
replace sensor.
Perform "Factory
Reset."
Replace key pad or
display or IOB.
Move glassware.
Fail: If cup is getting
power and wiring
has continuity
replace Detergent
cup.
Clean D-cup latch
Program warmer dry
temperature.
Tilt these articles
when loading so that
water drains off.
If labware is left in
the washer,
recondensation may
occur. Open latch
or program Auto
Unlock after
completion of cycle
to allow moisture-
laden air to escape.
Program longer
time.
Reconnect or
replace heater.
Reconnect or
replace blower.