Labconco SteamScrubber 4010010 Series User Manual page 97

Glassware washers
Table of Contents

Advertisement

Low Temp
No Water
Over Flow
Program Canceled
Power Fail
Program Pause.
Low Rinse Aid Cup
Defective fill sensor
Heater failure.
LT
Defective controller.
Defective
temperature probe.
Defective high
temperature cut out
T-Stat.
Leak during wash or
NW
rinse segment.
Drain Valve failure
during wash or rinse
segment.
OV
Water in base.
Program canceled
PC
by operator or Fatal
Alarm.
PF
Power Outage.
Door opened or
PP
program paused by
operator.
Rinse Aid in door low
RA
Rinse Aid Cup not
used.
Check Sensor
Screen: >125mm
when full.
Heater/s Test:
Check resistance
of heater for open
circuit
Check output from
IOB to heater.
Check electrical
connections.
Check resistance
of safety cut out
T-Stat.
Check for leaks
both under and
around washer.
Check for draining
during wash
segment.
Check for leaks
under wash tank.
No action
required.
No action
required.
No action
required.
97
Pass: Recalibrate or
replace sensor.
Contact Labconco
Product Service.
Fail: If heaters are
getting power and
wires have
conductivity, replace
heater/s
If open replace
heater.
Fail; Replace IOB.
Contact Labconco
Product Service.
Repair connections
If open circuit at
room temperature,
replace T-Stat.
Fix Leaks.
Replace drain valve.
Repair leaks under
wash tank or inlet
valves
No action required if
manually canceled.
Determine cause if
by another Fatal
Alarm.
Program will resume
at the beginning of
interrupted segment.
Restarting program
will resume where
paused.
Refill Rinse Aid in
door or dismiss
alarm.
Turn Rinse Aid Cup
dispense off in
Accessories.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents