Before you install the product, the installation site should be planned. Examine the
location where you intend to install it. You must be certain that the area is level and of
solid construction. In addition, a hot water source, a drain, a purified water source (if
applicable), and an electrical supply must be located near the installation site. A cold
water supply is required when the optional Cool Drain Water Kit is installed.

Location Requirements

The Glassware washer should be located in an area that is in close proximity to the
required utilities.

Clearance Requirements

If your glassware washer is to be mounted undercounter model, you must ensure the
undercounter opening is the proper size to accommodate the washer. The location
should provide adequate clearance for the door to lay down with access to load and
unload the racks.
See
Appendix B: Dimensions

Electrical Requirements

The product models have the following electrical requirements.
Catalog Number
4xxx01000
4xxx01010
4xxx01013
1 Electrical Requirements, 'V' = VAC (Voltage with alternating current), 'A' = Amperes
2 Operating with 208V, current draw approximately 21 A; Operating with 230V, current drawing approximately 19 A.
for overall product dimensions.
Heater
Typical Operating
Watts
Current (Amps)
1.2kW
16 A
2.0kW
11 A
6.0kW
21/19 A
11
Electrical Circuit Requirements
115 V, 60 Hz, 20 A
208-230 V, 50/60 Hz, 20 A
2
208/230/400 V, 50/60 Hz,
1
1 Phase
1 Phase
3 Phase