Calibration - Offsets

Adjust the sensor readings to match a known standard(s).
Current Reading (uncorrected)
Offset
= Value that will be added to the Current Reading (uncorrected). The
adjusted reading (uncorrected + offset) is the value reported on all screens
outside of the Sensor Calibration screen.
1. To adjust the calibration offset, touch
will display, which allows an offset value to be adjusted for the sensor.
2. Using a known standard as reference, change the offset to match the value
reported by the standard. For example, if the
uncorrected reading of 35
the offset to +1
When the
Cal. Date
Offset limits:
 Temperature = +/- 5
 Conductivity = +/- 100 µS
 Fill Level = +/- 5%
= raw sensor reading with no offset applied.
o
C, and the standard reference reports 36
o
C.
Offset
value is changed, the current date will be saved as the
o
C
Offset
field of the desired sensor. A pop-up
Sump / Water Temp
Figure 6-16
54
reports an
o
C, change
Last