Problem
Low Auto Rinse Aid
Conductivity
Sensor
Low Auto Detergent
Drain Fail
Alarm Code
Possible Cause
A1
Rinse Aid level low.
Float movement
restricted.
Bad wiring
connection.
Auto Rinse Aid not
used.
Conductivity sensor
CS
disconnected.
Bad conductivity
sensor.
Auto Detergent level
D1
low.
Float movement
restricted.
Bad wiring
connection.
Auto Detergent not
used.
Clogged Sump
DF
screen.
Diagnostic
Test
Check Rinse Aid
level.
Check Sensor
Screen for reading
and float function.
Check wire
connections.
Check wire
connections.
Check Sensor
Screen for a
conductivity value.
Check Auto
Detergent level.
Check Sensor
Screen for reading
and float function.
Check wire
connections.
Visual inspection.
94
Possible
Corrective
Action
Refill or replace
Rinse Aid bottle.
Clean float on dip
tube.
Correct wiring.
Turn Auto Rinse Aid
dispense off in
Accessories.
Dismiss alarm.
Correct wiring.
If no reading and
wiring connected
replace conductivity
sensor.
Refill or replace
Detergent bottle.
Clean float on dip
tube.
Correct wiring.
Turn Auto Detergent
dispense off in
Accessories.
Dismiss alarm,
Clean sump screen.

Hide quick links: