Alarms; Completing The Wash Cycle - Labconco SteamScrubber 4010010 Series User Manual

Glassware washers
Table of Contents

Advertisement

Alarms

Throughout a program cycle there are several abnormal events that could occur. The
washer is programmed to monitor these occurrences and provide an alarm to alert the
user. Information is displayed to identify the alarm situation, which may be Fatal or
Non-Fatal to the running program. See Table 7-1 for a list of all Alarm Codes and their
description. Fatal alarms during a cycle will drain the washer, end the program and
display a red
X
on the segment in which the error occurred. Non-Fatal alarms will allow
the program to proceed. The display will show a yellow check mark at any complete
segment in which a non-fatal alarm occurred. Touching the red
display alarm details and the time of occurrence.
ALARM
ALARM DESCRIPTION
CODE
A1
LOW AUTO RINSE AID 1
D1
LOW DETERGENT 1
DF
DRAIN FAIL
DL
DOOR LATCH FAIL
DO
DOOR OPEN
FF
FILL FAIL
FI
FIRMWARE
HW
HIGH WATER
LT
LOW TEMP
NW
NO WATER
OV
OVER FLOW
PC
PROGRAM CANCELED
PF
POWER FAIL
PP
PROGRAM PAUSED
RA
RINSE AID LOW D-CUP
SS
STEP SKIPPED
TS
TEMP SENSOR
*
Only on models with optional Auto Dispense or Rinse-Aid Dispense accessory installed.

Completing the Wash Cycle

The wash cycle can be terminated at any time by pressing the
confirming the cancelation. The washer will drain and end the cycle. The progress bar
will display a red
X
If the cycle is completed with no interruptions or alarms, the progress bar will display
blue checks at each segment, and a Program Complete pop-up will display until
acknowledged.
If Auto-Unlock was enabled in the program, once the internal temperature is below 50
the door latch will open to allow the washer to vent and assist drying.
Pressure Sensor did not reach 0 during drain
at the segment in which the program was cancelled.
CONDITION
Low Rinse Aid in Auto Dispense *
Low Detergent in Auto Dispense *
Door unlatched during cycle
Door not fully closed
Low Water after allotted fill time expires
Firmware Unresponsive
Fill Level too high during segment
Temperature Set Point not achieved
No or Low Water during segment
Water detected in base
User initiated Program Cancel
Power Loss occurred
User initiated Program Pause
Low Rinse Aid in Detergent Dispense
User initiated Segment Skip
Temp Sensor error or disconnected
81
X
or yellow check will
Table 7-1
RESPONSE
NON-FATAL
NON-FATAL
FATAL
FATAL
FATAL
FATAL
NON-FATAL
FATAL
FATAL
FATAL
FATAL
FATAL
NON-FATAL
NON-FATAL
NON-FATAL
NON-FATAL
FATAL
Cancel
Icon and
o
C

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents