Calibration - Setpoints - Labconco SteamScrubber 4010010 Series User Manual

Glassware washers
Table of Contents

Advertisement

Calibration - Setpoints

Adjust the maximum program segment temperature.
Setpoint
= maximum programmable temperature in any segment.
1. To adjust the Setpoint, touch
6-17. A pop-up will display, which allows the maximum set point to be set.
2. The pop-up provides two choices: 88
3. If the washer is installed at an elevation above 2000 meters (6561 feet), change
the maximum temperature set point to 88
When the
Last Cal. Date
Setpoint
Setpoint
value is changed, the current date will be saved as the
field of the desired value. Reference Fig.
o
o
C or 93
C.
o
C. Touch [OK].
Figure 6-17