 Locate the holes on the bottom of the washer where the four skid-fastening bolts
were previously attached. See Fig. 4-2.
 Thread the four leveling feet into the holes until the overall distance from the
bottom of the foot to the top of the washer is approximately the height required to
fit in the opening on under counter models. The height of Freestanding models
typically match the counter top.
 Carefully move the washer into an upright position. Be careful to distribute the
weight evenly on the leveling feet when setting the washer upright.
 If installing a freestanding unit remove the back panel. Move the washer as close
as possible to its final location but maintain access to the water connections and
drain hose in the back. The water supply hoses will connect at the back of the
washer. The wiring for the electrical connection will run from the rear of the unit
through to the electrical box located behind the front toe kick panels.
 Adjust the leveling feet to level the washer and raise the unit to the approximate
desired finished height. A final adjustment may be required after the washer is
placed in its final position.
 Additional accessories, including Automatic Detergent Dispense, Rinse Aid
Dispense or Cold Water Drain Kits, should be installed at this time. Refer to the
assembly instructions included with each kit. Individual kit instructions are also
included on the User's Manual USB Thumb Drive (shipped with the washer) or
online at Labconco.com.

Hot Water Connection

A hot water supply line must be connected to the Glassware Washer.
 To prevent valve clogging, thoroughly flush the hot water lines from the supply
source prior to connecting the washer to the water lines. The hot water inlet
valve is located on the back of the washer base. The back panel of freestanding
models must be removed to access the valve.
 The washer's hot water inlet valve is equipped with a male ¾-11 ½ GHT fitting.
Connect the hot water supply coming from the customer-supplied shut off valve
using an appliance style braided hose with ¾-11 ½ GHT fitting and ¼" minimum
inside diameter. See Fig. 4-3.
 On undercounter installations, an extra 3' of service loop is recommended for the
supply hose to allow the washer to be slid in or out of its operating position.