Fill And Heat Specifications; Environmental Conditions - Labconco SteamScrubber 4010010 Series User Manual

Glassware washers
Table of Contents

Advertisement

Fill and Heat Specifications

 Minimum fill water temperature:
 Water consumption per fill:
 Minimum tap feed water pressure:
 Maximum tap feed water pressure:
 Minimum tap feed water flow rate
 Minimum feed purified water pressure: 0 psi (0 kPa)
 Maximum feed purified water pressure: 120 psi (827 kPa)
 Minimum pure water flow rate
 Heat rejected from washer* 1200W:
 Heat rejected from washer* 2000W:
 Heat rejected from washer* 6000W:
 Water heating rate** 1200W 115V:
 Water heating rate^ 230V 2000W:
 Water heating rate^ 208V 6000W :
*Wash Pump and Heater(s) On
**1200W Models - Heat Rate from 50
^2000W and 6000W Models - Heat Rate from 50

Environmental Conditions

Indoor use only
Ambient temperature range: 41° to 104°F (5° to 40°C)
Maximum altitude: 10,000 feet (3,048 meters).
Maximum relative humidity: 80% for temperatures up to 88°F (31°C), decreasing
linearly to 50% relative humidity at 104°F (40°C)
Main supply voltage fluctuations not to exceed ±10% of the nominal voltage
Transient overvoltages according to Installation Categories II (Overvoltage
Categories per IEC 1010). Temporary voltage spikes on the AC input line that may
be as high as 1500V for 115V models and 2500V for 230V models are allowed
Used in an environment of Pollution degrees 2 (i.e., where normally only non-
conductive atmospheres are present). Occasionally, however, a temporary
conductivity caused by condensation must be expected in accordance with IEC 664
120°F (49°C) recommended
3.4 gallons (13 L)
20 psi (138 kPa)
120 psi (827 kPa)
1.25 gallons (4.7 liters) per minute
.9 gallons (3.4 liters) per minute
< 68 BTU/minute
< 114 BTU/minute
< 341 BTU/minute
1.2°F/minute (0.7°C/minute)
3.1°F/minute (1.7°C/minute)
5.8°F/minute (3.2°C/minute)
o
C to 82
o
C
o
C to 93
o
C
108