Recommended separation distances
from wireless communication devices
WARNING
Risk due to electromagnetic disturbance
Wireless communication devices (e.g., cellu-
lar phones) and medical electrical equipment
(e.g., defibrillators, electrosurgical devices)
emit electromagnetic radiation. When such
devices are operated too close to this device
or its cables, the functional integrity of this
device may be compromised by electromag-
netic disturbances. As a result, the patient
could be put at risk.
Maintain a distance of at least 0.3 m (1.0 ft)
between this device and wireless communica-
tion devices to ensure that the essential per-
formance of this device is fulfilled.
Maintain an adequate distance between this
device and other medical electrical equip-
ment.
Instructions for use Oxylog 3000 plus SW 1.n
Technical data
153