Activating options

The activation codes for options can be entered.
The activated options are then displayed.
For the possible options, refer to "List of
accessories" on page 165.

Checking keys and control knobs

The operating elements on the front panel are
displayed schematically on the screen.
– Display = screen
– B = buttons
To test the control knobs, set the following values:
– VT to 500 mL
– RR to 20/min
– Pmax to 40 mbar
– FiO
to 40 %
2
These settings are displayed on the screen.
To test the buttons:
Instructions for use Oxylog 3000 plus SW 1.n
1 Briefly press the corresponding button.
The associated letter on the screen changes
from "B" to "X". If the button has an indicator, it
will be illuminated by the device. If there are
buttons without an indicator, the yellow warning
indicator will light up on the device.
NOTE
If you press the
key for longer than 3 seconds,
the ventilator switches OFF.
The function of the rotary knob is not included in the
test.
Checking the loudspeaker, buzzer, LEDs
and display
To test the loudspeaker, buzzer, all LEDs and the

display:

1 Select the required test.
2 Confirm the selection by pressing the rotary
knob. The requested function is tested by the
device.
To test the screen display (Test display).
3 Turn the rotary knob; various test cards are
displayed.
The selected test remains active until the rotary
knob is pressed again.