Advertisement

Determining the approximate pneumatic operating time
Example for supply of O
– Cylinder pressure measured on the manometer
of the pressure reducer: 20000 kPa
(200 bar)
– Liquid capacity of the O
Supply of O
:
2
2.1 L × 20000 kPa = approx. 420 L at ambient
pressure.
Example for pneumatic operation time:
– VC-CMV mode; frequency: 10 breaths/min,
VT = 0.53 L, O
= 100 %
2
– Minute volume = 10 breaths/min x 0.53 L =
5.3 L/min
O
supply [L]
Operation time =
2
(MV +0.5*) [L/min]
* Average gas consumption of the ventilator:
0.5 L/min
420
Operation time =
5.8
The pneumatic operation time increases when the
Oxylog 3000 plus operates with an O
concentration of less than 100 % O
is drawn into the device.
The amount of gas from the high-pressure supply,
which is currently being consumed, is indicated by
the Oxylog 3000 plus in the lower left section of the
information window in L/min. This display is
overwritten when a higher priority message is
activated.
Instructions for use Oxylog 3000 plus SW 1.n
:
2
cylinder: 2.1 L
2
= approx. 72 minutes
2
, as ambient air
2
Example:
A
A O
consumption = 5.3 L/min
2
Preparation
53