Configuration
Displaying power supply data
The parameters of the rechargeable battery and
the status of the external power supply are
displayed.
Display (example):
Advance to the second page:
Select line Page (Page), confirm and turn the
rotary knob.
Display (example):
102
Checking the safety valve
For safety inspection, a flow can be generated to
test the safety valve.
Displaying the technical logbook
Any technical errors and/or special occurrences,
such as activation of a software option, completion
of the device check and device calibration, are
listed in chronological sequence.
The entries in the logbook cannot be deleted and
are retained when the device is switched on and off
or after a power supply failure.
A maximum of 570 logbook entries can be
displayed on the screen. When the capacity of the
logbook is reached, the oldest entries are
overwritten.
Display (example):
Instructions for use Oxylog 3000 plus SW 1.n