For Your Safety and that of Your Patients
Product-specific safety information
WARNING
Ventilation monitoring is mandatory at all
times! Whenever a patient is connected to the
ventilator, constant attention by qualified
medical staff is required in order to provide
immediate corrective action in case of a
malfunction.
The user must not solely rely on the built-in
monitoring of the ventilator and must always
assume full responsibility for proper
ventilation and patient safety in all situations.
WARNING
Keep an manual resuscitator at the ready
If a malfunction is detected in the ventilator
and its life-support functions can no longer be
guaranteed (such as in case of a power supply
failure or interruption in the compressed gas
supply), ventilation must be started without
delay with an independent ventilator (manual
resuscitator) – using PEEP and/or increased
inspired O
concentration as necessary.
2
WARNING
Risk of CO
rebreathing
2
To ensure proper ventilation when setting the
ventilation parameters, the total dead space
volume of the breathing circuit must be
considered. This applies particularly when
using low tidal volumes. Observe for signs of
rebreathing.
WARNING
Risk of malfunction
Unauthorized modification of the medical
device will result in malfunction.
This medical device must not be modified
without the permission of Dräger.
12
CAUTION
An etCO
value by itself is insufficient as a basis
2
for medical decisions.
Installing accessories
CAUTION
Installations on the Oxylog 3000 plus must be
done in accordance with these instructions for
use. Make sure that the connections are securely
fitted onto the basic unit system.
Strictly follow the assembly instructions and
instructions for use.
Only one copy of instructions for use
included
CAUTION
Only one copy of the instructions for use is
included per device, and it must therefore be kept
in an accessible location for users.