Semicritical Medical Devices - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Cleaning, Disinfection and Sterilization
Class of active
Product name
ingredient
Chlorine-
Actichlor
releasing
Klorsept
agents
BruTab 6S
Oxygen-
Descogen
releasing
Liquid
agents
Descogen
Liquid
Dismozon
Dismozon
Oxycide
Perform
Virkon
Quarternary
Mikrozid
ammonium
sensitive liquid
compounds
Mikrozid
sensitive wipes
Aldehydes
Buraton
1) Virucidal against enveloped viruses
Dräger recommends using a surface disinfectant
from the list above. Other disinfectants are used at
own risk.
Dräger points out that oxygen- and chlorine-releas-
ing agents may cause color change in some mate-
rials. This is not an indication that the product is not
functioning correctly.
At the time of validation, the disinfectants listed
above showed material compatibility.
The manufacturers of the surface disinfectants
have verified at least the following spectra of
activity:
– Bactericidal
– Yeasticidal
– Virucidal or virucidal against enveloped viruses
Observe the specifications of the surface
disinfectant manufacturers.
126
Manufacturer
®
plus
Ecolab
®
17
Medentech
Brulin
®
Antiseptics
®
®
plus Bode chemie
®
pur
®
Ecolab USA
®
Schülke & Mayr
®
DuPont
®
Schülke & Mayr
1)
®
1)
®
10 F
Schülke & Mayr
Carry out manual cleaning and disinfection
 Remove dirt immediately with a wipe soaked in
disinfectant.
WARNING
Penetrating liquid may cause malfunction of
or damage to the device, which may endanger
the patient. Only wipe-disinfect items and
make sure no liquids penetrate into the
device.
1 Carry out surface disinfection.
2 After contact time, remove disinfectant
residues.

Semicritical medical devices

Testing of procedures and agents
The cleaning and disinfection of medical devices
has been tested with the following procedures and
agents. The following agents showed good material
compatibility and effectiveness at the time of the
test:
Manual cleaning:
– For the breathing valve and hoses:
Neodisher LM2 by Dr. Weigert
(concentration: 2 %)
– Flow sensor:
Sekusept Powder classic by Ecolab
(concentration: 4 %)
Manual disinfection:
– Korsolex extra by Bode Chemie
(concentration: 3 %)
Machine cleaning and disinfection:
Neodisher Forte MediClean from Dr. Weigert
(for thermal disinfection at 93 °C (199 °F) for
10 minutes)
Instructions for use Oxylog 3000 plus SW 1.n

Advertisement

Table of Contents
loading