Monitoring
Expiratory minute volume MVe
Alarm, upper alarm limit
Alarm, lower alarm limit
Airway pressure, Paw
Alarm, upper alarm limit
Alarm, lower alarm limit
Apnea alarm time Tapn
Alarm
Apnea
Alarm
Rate, RRsp
Alarm, upper alarm limit
Leakage
Alarm
Instructions for use Oxylog 3000 plus SW 1.n
Range of settings
Range of settings
Range of settings
Range of settings
Range of settings
When the upper alarm limit has been exceeded.
2 to 41 L/min
When the lower alarm limit has been violated.
0.5 to 40 L/min
When value Pmax is exceeded.
20 to 60 mbar
When the pressure difference between inspiratory
and expiratory phases is less than 5 mbar for at
least 20 seconds.
Or
When the set pressure value is not attained within
10 seconds.
When respiratory activity is no longer detected.
15 to 60 s
When no respiratory phase change has been
detected for >12 seconds and Tapn is off.
> 25 seconds above the set value
10 to 100/min
Only with NIV off in VC modes:
When VTe <60 % of VTi over the last 25 seconds or
8 respiratory cycles
Technical data
145