Monitoring; Expiratory Minute Volume Mve; Airway Pressure, Paw; Apnea Alarm Time Tapn - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Monitoring

Expiratory minute volume MVe

Alarm, upper alarm limit
Alarm, lower alarm limit

Airway pressure, Paw

Alarm, upper alarm limit
Alarm, lower alarm limit

Apnea alarm time Tapn

Alarm

Apnea

Alarm

Rate, RRsp

Alarm, upper alarm limit

Leakage

Alarm
Instructions for use Oxylog 3000 plus SW 1.n
Range of settings
Range of settings
Range of settings
Range of settings
Range of settings
When the upper alarm limit has been exceeded.
2 to 41 L/min
When the lower alarm limit has been violated.
0.5 to 40 L/min
When value Pmax is exceeded.
20 to 60 mbar
When the pressure difference between inspiratory
and expiratory phases is less than 5 mbar for at
least 20 seconds.
Or
When the set pressure value is not attained within
10 seconds.
When respiratory activity is no longer detected.
15 to 60 s
When no respiratory phase change has been
detected for >12 seconds and Tapn is off.
> 25 seconds above the set value
10 to 100/min
Only with NIV off in VC modes:
When VTe <60 % of VTi over the last 25 seconds or
8 respiratory cycles
Technical data
145