Alarm
Device failure
!!!
!!!
Display inop
!!
etCO
high
2
!!
etCO
low
2
!!
Flow meas-
urement inop
!!
High respira-
tory rate
!!
Int. battery
charging inop
Instructions for use Oxylog 3000 plus SW 1.n
Cause
Technical defect.
Technical defect.
The upper alarm limit for endtidal
CO2 concentration has been
exceeded.
The lower alarm limit for endtidal
CO2 concentration has been
exceeded.
Measurement lines for flow
measurement kinked, discon-
nected or leaking.
Faulty flow sensor.
Technical defect.
Patient breathes at a high
spontaneous rate.
The internal battery is not being
charged due to a battery failure.
The internal battery is not being
charged due to a device malfunc-
tion.
Remedy
Disconnect the patient from the
device and continue ventilation
without delay using another
ventilator.
After pressing Alarm Reset the
device will directly display fur-
ther failure information in the
"Customer Service Mode"
menu. Call DrägerService.
Disconnect the patient from the
device and continue ventilation
without delay using another
ventilator. Call
DrägerService.
Check patient condition.
Check alarm limits.
Adjust the alarm limit, if
necessary.
Check patient condition.
Check alarm limits.
Adjust the alarm limit, if
necessary.
Ensure flow measuring lines are
connected correctly.
Replace the flow sensor.
Disconnect the device from the
patient and continue ventilation
without delay using another
ventilator. Call DrägerService.
Check patient condition, check
ventilation pattern, correct alarm
limit RRsp if necessary.
Change internal battery.
Call DrägerService.
Continuous ventilation with this
device is only possible with an
external power source. Call
DrägerService.
Problem solving