Disassembly; Cuvette; For Adults - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Cleaning, Disinfection and Sterilization

Disassembly

Disassembling the CO

cuvette

B
A
1 Disconnect the CO
device.
2 Remove the CO
sensor (B) from the CO
2
cuvette (C).
3 Remove the CO
cuvette (C) from the
2
flow sensor (D).
4 Remove the angled connector (A)
from the cuvette (C).
122
sensor and CO
2
D
C
sensor connector from the
2
Disassembling the reusable breathing
2
circuit for adults
2
1 Disconnect the breathing hose (A) from the
expiratory port.
2 Detach the flow measuring lines (B) from the
connectors.
Instructions for use Oxylog 3000 plus SW 1.n
B
A

Advertisement

Table of Contents
loading

Table of Contents