Assembly
Connecting the CO
B
A
1 Disconnect the angled connector (A)
from the flow sensor (D).
2 Attach the cuvette (C) to the flow sensor (D),
with the cuvette windows facing the side.
3 Attach the angled connector (A) to the
cuvette (C).
4 Push the CO
sensor (B) onto the cuvette (C),
2
with the cable towards the device.
5 Connect the CO
sensor to the connector of the
2
Oxylog 3000 plus. For the connector location,
refer to the section "Side view, right" on
page 19.
6 Insert the CO
sensor cable in the cable clips on
2
the hose.
Alternatively, connect the cuvette (C) directly to the
patient side of the angled connector (A), without
disconnecting the angled connector from the flow
sensor (D).
The CO
sensor cable can be extended with a
2
maximum of one extension cable. Refer to the "List
of accessories" on page 165.
48
sensor and the cuvette
2
D
C
For information on CO
testing before ventilation, see page 58. For
information on CO
zero calibration and filter
2
testing during ventilation, see page 91. For
information on CO
measurement and cuvette type
2
selection, see page 91. For information on CO
configuration in Customer Service Mode, see
page 103.
Instructions for use Oxylog 3000 plus SW 1.n
zero calibration and filter
2
2