3 Disconnect the flow sensor (A) from the
breathing valve.
4 Carefully detach the flow measuring lines (B)
from the flow sensor.
Pull the flow measuring lines straight off the
connectors.
CAUTION
Do not twist or use force when disconnecting
the flow measuring lines from the flow sensor
nozzles.
Otherwise the flow sensor may be damaged.
5 Detach the angled connector (C) from the flow
sensor.
6 Detach the breathing hose (D) from the
breathing valve.
Instructions for use Oxylog 3000 plus SW 1.n
D
B
B
Cleaning, Disinfection and Sterilization
Disassembling the breathing valve
A
B
7 Turn the cover (A)
about 90° counterclockwise to unlock it.
8 Remove the silicone diaphragm (B).
WARNING
Risk of CO
rebreathing
2
The rubber disk (A) in the housing must not be
removed, damaged or bent, otherwise the
valve will not work properly and will endanger
the patient.