– Insp.term.%PIF
Inspiration termination criteria of pressure
supported strokes, as percentage of the peak
inspiratory flow (PIF).
– Hose type
The selected hose type must match the hose
type in use. Otherwise a correct volume
measurement cannot be guaranteed.
– CO
cuvette type (optional)
2
NOTE
The cuvette windows of the reusable cuvette and
disposable cuvette have different optical
properties. Therefore, the correct cuvette type
must be selected in the settings menu. Otherwise
the zero point will be shifted by as much as
±8 mmHg CO
.
2
Instructions for use Oxylog 3000 plus SW 1.n

VC-SIMV/PS

The following parameters can be adjusted on-
screen in addition to the parameters of ventilation
mode VC-SIMV:
– Setting on page 1: Pressure Support
above PEEP
– Setting on page 2: Pressure rise time Slope
(condition:
Psupp > 0 mbar)
Flat slope = slow pressure rise
Medium slope = medium pressure rise
Steep slope = rapid pressure rise

Setting AutoFlow (optional)

The AutoFlow-AF function can be set for VC-SIMV
and SIMV/PS.
When AutoFlow is switched on, the setting Tplat is
no longer valid, and Slope must be set.
For more information on AutoFlow, refer to
"AutoFlow" on page 160.
Operation
Psupp
69