Data Communication (Option); Operating Data; Power Supply - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Technical data

Data communication (option)

Exported data
Baud rate
Data bits
Parity
Stop bit

Operating data

Power supply

Power supply
Power consumption
Battery type
Charging time
146
Oxylog 3000 plus
Input voltage
Power supplies (AC/DC
power pack and DC/DC
converter) are specified as
part of the
Oxylog 3000 plus.
With battery charging
Operating time with a new
and fully charged internal
battery without mains
power supply
– in typical ventilation
without CO
sensor,
2
reduced display
brightness
– in typical ventilation
Measured values
Curves
Alarms
Alarms settings
User settings
For the data communication protocol contact your
local DrägerService.
19200
8
Even
1
24 V ±6 VDC
maximum 2.4 A at 19 VDC
Approx. 9.5 hours
Approx. 7.5 hours
Lithium ion battery
Approx. 4 hours
The specified charging time applies when recharg-
ing the battery to a level of approx. 90 % after it has
been depleted. Charging of a completely depleted
battery is only possible when the ventilator is
switched OFF.
Instructions for use Oxylog 3000 plus SW 1.n