OXYLOG 3000 PLUS:
QUICK START GUIDE
1
How do I connect the
ventilator? (Image A)
• Insert power supply for mains power
(unless internal battery charged). Attach
patient breathing circuit
, flow sensor
and oxygen supply
.
• Connect high-pressure oxygen supply at
(via cylinder/pipeline with Schrader valve)
Flow sensor measurement port
Connection point for breathing circuit
Oxygen supply connection
Connection for power supply
Connection for CO2 sensor
Connection point for data communication cable
3
Navigation system for
setting controls (Image C)
Please see default settings labelled on the device. See labelled diagram for knob/buttons.
Display screen
Alarms
Settings
Ventilation mode buttons: Spn CPAP,
VC-CMV/VC-AC, VC-SIMV, PC-BIPAP.
Red/yellow alarm indicator
Alarm mute (for 2 minutes)
Alarm reset (alarm acknowledge)
100% 0
2
This quick start guide has been developed by the NHS and should be used in conjunction with the manufacturers Instructions for Use
CONNECT AND SET UP
A
6
1
2
5
4
3
SET THE CONTROLS
Inspiratory hold (manual inspiration)
Standby
11 Rotatory knob (changing
11
and confirming settings)
12 Power source (battery or mains)
12
13 Curves (to change between the pressure,
13
flow or CO2 curves on display)
14 Values (to change screen pages in
14
the measured values window)
System Overview
Adult hose system, reusable
2
How do I set up the breathing system
and the HME filter? (Image B)
• Set-up using disposable hose system
,
• To breathing hose
attach breathing valve
. Flow sensor
attaches to breathing
valve
.
• The flow sensor
has an integrated flow
and pressure sensor tubing
attaches to right-side panel of device (A
A Angled connector
B Flow sensor
• HME
to be connected to angled piece
C Breathing valve
If HME used, this adds resistance to system,
D Breathing hose
therefore need to set 'HME correction' when
E Flow and pressure measuring hoses
prompted on display.
14
13
Control knob:
tidal volume
Control knob:
respiratory rate
20
Control knob:
pressure max
(Pmax)
TRANSPORT VENTILATOR
Adult hose system, disposable
B
2
B
1
A
, which
1
).
A Angled connector
2
B Flow sensor
.
3
C Breathing valve
4
D Breathing hose
5
E Flow and pressure measuring hoses
6
HME filter addition to patient circuit
Instructions for Use
Oxylog 3000 plus
C
2
1
Instructions for Use Oxylog 3000 plus SW 1.n
11
Control knob: 0
2
Emergency and Transport Ventilator
concentration (Fi0
WARNING
)
2
To properly use this medical device,
Software 1.n
read and comply with these
Instructions for Use.
4 5
3
D
C
E
6
3
4
5
6
7
8
9
10