Operation
Ending apnea ventilation
 Press the Alarm Reset key.
The ventilator is again ventilating with the previous
settings in ventilation mode Spn-CPAP.
Deactivating apnea ventilation
 Set Tapn to OFF.
NOTE
Apnea ventilation can only be activated in
ventilation mode Spn-CPAP without NIV.
The minimum minute volume required by the
patient must be monitored via the lower alarm limit
MVe
.

Spn-CPAP/PS

If
Psupp is set to a value above 0 mbar, the
following parameters can be adjusted on-screen in
addition to those for Spn-CPAP:
– Trigger sensitivity
Successful patient triggering is indicated by an
asterisk (*) on the left side of the curve window.
– Pressure rise time Slope (effective for Pressure
Support
Psupp)
74

Cardio-pulmonary Resuscitation (CPR)

Refer to the "Cardiopulmonary resuscitation
(CPR)" on page 67.
Instructions for use Oxylog 3000 plus SW 1.n