Nominal flow
Dead space including flow sensor, but
excluding accessories such as filter, HMEs
and CO
cuvette
2
Dead space of CO
2
cuvette
Supplementary functions
Patient connection
1) At a supply pressure >350 kPa.
The maximum inspiratory flow is reduced to 80 L/min at supply pressures <350 kPa and to 39 L/min at supply pressures
<280 kPa.
2) When using an accessory with a female connector, add 2 mL to the breathing circuit dead space.
Instructions for use Oxylog 3000 plus SW 1.n
Disposable breathing
circuit for adults, 1.5 m
Disposable breathing
circuit for adults, 3.0 m
Reusable breathing
circuit for adults, 1.5 m
Reusable breathing
circuit for adults, 3.0 m
Disposable breathing
circuit for pediatric
patients
Demand valve
Relief valve
Inspiratory 21 L/min at 2 mbar
Expiratory 10 L/min at 2 mbar
Inspiratory 20 L/min at 2 mbar
Expiratory 10 L/min at 2 mbar
Inspiratory 22 L/min at 2 mbar
Expiratory 16 L/min at 2 mbar
Inspiratory 21 L/min at 2 mbar
Expiratory 15 L/min at 2 mbar
Inspiratory 12 L/min at 2 mbar
Expiratory 7 L/min at 2 mbar
2)
Approx. 35 mL
(reusable breathing circuit for adults)
2)
Approx. 30 mL
(disposable breathing circuit for adults)
2)
Approx. 15 mL
(Disposable breathing circuit for pediatric patients)
Approx. 4 mL (CO
cuvette)
2
Approx. 1.5 mL (CO
cuvette for pediatric patients))
2
Opens the breathing system upon failure of the gas
supply, permits spontaneous breathing with ambient
air.
Opens the breathing system in case of device
malfunction at approximately 80 mbar.
22 mm ISO conical connector
Technical data
143