Problem solving

Messages in the information field

The numeric values below are examples.
Message
Confirm PEEP above 10 mbar ?
Gas consumption = 10 L/min
(Battery capacity)
REUS = Adult reusable hose
DISP = Adult disposable hose
PAED = Pediatric disposable
hose
Select and confirm hose with
rotary knob.
REUS = Reusable Cuvette
DISP = Disposable Cuvette
Press Rotary Knob for
Autoset / Exit with Alarm Reset
Remove sensor from cuvette
then press Rotary Knob
Place sensor on reference
filter, press rotary knob
Pinsp>=PEEP + 3 mbar
118
1)
Unit
Cause
PEEP >10 mbar has been set
but not confirmed.
Default display for current
gas consumption.
Default display for current
battery capacity.
Explanation of abbreviation,
when selecting the hose type
in the Settings window.
Breathing hose type has not
been confirmed.
Explanation of abbreviation,
when selecting the CO
cuvette type in the Settings
window.
Activation of Alarms: Auto-
set, refer to "Setting alarm
limits" on page 87.
Activation of CO
bration, refer to "CO
urement (optional)" on
page 91.
Activation of CO
check, refer to "CO
urement (optional)" on
page 91.
PEEP + 3 mbar > Pinsp
Explanation/Remedy
The required setting of
PEEP >10 mbar is only pos-
sible when confirmed via the
rotary knob.
Select and confirm.
2
After pressing the rotary
knob the new alarm limits
will be set.
zero cali-
After removal of the CO
2
meas-
sensor from the cuvette and
2
pressing the rotary knob, the
zero calibration will proceed.
Cancel by pressing Alarm
Reset.
filter
After mounting the test filter
2
meas-
and pressing the rotary
2
knob, the filter check will
start. Cancel by pressing
Alarm Reset.
Set Pinsp > PEEP + 3 mbar.
Instructions for use Oxylog 3000 plus SW 1.n