The Oxylog 3000 plus displays the alarm message
Filter check failed (Filter check failed). The
measured CO
value is outside the permissible
2
tolerance range.
 Check whether the sensor (A) or test filter (B) is
soiled and clean them if necessary. Repeat the
CO
filter check.
2
If the check was still not successful:
 Perform the CO
gas check.
2
Performing the CO
gas check
2
Perform the CO
gas check if the required CO
2
value is not reached during the CO
NOTE
Before the CO
gas check can be performed, the
2
CO
zero calibration must be completed
2
successfully. Otherwise the CO
be outside of the tolerance range.
NOTE
For the CO
gas check use only a test gas
2
composed of CO
and N
2
Otherwise values other than ±0.5 Vol% CO
be displayed.
1 Use the reusable cuvette from the calibration
set.
D
B
2 Connect the test gas cylinder (A) and the
cuvette (B) of the calibration set to the hose (C).
3 Fit the CO
sensor (D) on the cuvette (B) of the
2
calibration set and connect the CO
the ventilator.
Instructions for use Oxylog 3000 plus SW 1.n
2
filter check.
2
gas check may
2
.
2
may
2
E
C
A
sensor to
2
4 Open the page CO
select the line Filter/gas check (Filter/gas
check) and then the line Gas check (Gas
check) and confirm. The display now shows
Turn on test gas (Turn on test gas).
5 Read the CO
concentration of the test gas
2
from the test gas cylinder (A).
6 Open the test gas cylinder (E) and set the test
gas flow to 0.1 L/min. Select CONTINUE
(CONTINUE) and confirm with the rotary knob.
7 The Oxylog 3000 plus displays the CO
concentration Measured CO
CO
value).
2
8 About 1 minute after setting the test gas flow,
check if the displayed CO
CO
concentration specified on the test gas
2
cylinder (tolerance: ±0.2 Vol%).
9 Close the test gas cylinder.
If the measured CO
value is outside the permitted
2
tolerance, the CO
sensor must be recalibrated
2
with test gas.
CO
calibration with test gas
2
The CO
sensor must be calibrated if the required
2
CO
value is not reached during the CO
2
check.
NOTE
Before the CO
calibration with test gas can be
2
performed, the CO
zero calibration must be
2
completed successfully. Otherwise the CO
calibration with test gas may be outside of the
tolerance range.
NOTE
For the calibration use only a test gas composed
of CO
and N
.
2
2
Otherwise values other than ±0.5 Vol% CO
be displayed.
Configuration
sensor (CO
sensor),
2
2
2
value (Measured
2
value matches the
2
gas
2
2
2
may
105