Teledyne AutoInjector Module Installation And Operation Manual page 60

Automation modules
Table of Contents

Advertisement

Automation
Section 4 The Column Selector Module
Figure 4-24 Check valve connected to the waste line of the
AutoInjector Module
Figure 4-25 Sample tube fitting attached to the EZ Prep
inject valve
4-14
18. Locate the supplied sample probe tubing and fitting pack-
P
age (
/
60-5234-657).
N
19. Unscrew the attached plastic ferrule in this fitting package
until the plastic ferrule is almost completely detached from
the steel fitting.
20. Attach the steel fitting in this package to the EZ Prep
inject valve waste port. This steel fitting should be hand
tight.
21. Turn the plastic ferrule until it is attached hand tight to
the valve.
22. Turn the steel fitting
23. Remove the plastic ferrule while leaving the fitting
attached.
1
/
turn tighter.
2