Automation
Section 3 The AutoSampler Module
3-22
Table 3-1 Queue Tab Columns
Column
=
Sample Name
Column (drop down menu)
Method
Sample Position
Sample Volume
Description
Allows the user to remove a single
row of an existing queue, clear all
rows of the queue, or R
N
UN
Allows the user to name each sam-
ple. Naming is optional. If left blank,
the sample is named based on the
date and time the separation was
started.
If a Column Selector Module is
installed, this menu allows you to
select a separation device and asso-
ciated method for use.
If a Column Selector Module is not
installed, this menu allows you to
select different methods that may be
associated with the configured and
installed separation device.
Once a column has been selected,
or if there is only one column
defined on the system, the default
method created for that column is
automatically loaded. To change this
method, touch the method name
and a list of methods associated
with the column will appear. These
methods will be listed in the order
they were created. Alternatively,
select C
C
USTOMIZE
URRENT
or B
to populate this field.
ROWSE
Select C
C
USTOMIZE
URRENT
to create a modified version of the
method for use in the current queue.
To use this modified method, exit the
method editor and select S
AVE
prompted. A new method named
"Temporary 1" will be created. This
method will be used for the current
sample and will be discarded after
the sample is complete. All method
parameters are saved with each
separation. If you would like to save
the modified method for later use,
select S
A
as you exit the
AVE
S
method editor and create a unique
name for the new method.
This field indicates where the sam-
ple for this separation is located. If
the sample size is too large for a sin-
gle vial, create a second position in
the queue and access the second
vial.
Used in conjunction with the N
I
column to deter-
BER OF
NJECTIONS
mine the size of each injection.
.
EXT
M
ETHOD
M
ETHOD
when
-
UM