Teledyne AutoInjector Module Installation And Operation Manual page 16

Automation modules
Table of Contents

Advertisement

Automation
Section 2 The AutoInjector Module
Figure 2-9 Waste line connected to the right AutoInjector
Module port
Figure 2-10 Check valve connected to the waste line of the
AutoInjector Module
2-6
17. Connect the check valve removed in Step 4 to the waste
line from the AutoInjector Module. Ensure that the arrow
on the check valve points away from the AutoInjector Mod-
ule as shown in Figure 2-10.
18. Locate the supplied sample probe tubing and fitting pack-
P
age (
/
60-5234-657).
N
19. Unscrew the attached plastic ferrule in this fitting package
until the plastic ferrule is almost completely detached from
the steel fitting.
20. Attach the steel fitting in this package to the EZ Prep
inject valve waste port. This steel fitting should be hand
tight.