Figure 3-15 Illustration of the proper probe cable route
through the rotor
Note
Be sure to route the probe cable through the proper receptacle
on the rotor. The probe cable should follow the groove as
shown in Figure 3-15.
19. Loosen the thumbscrew on the rotor to allow the cable to
pass through the clamp. Extend about 2 mm of cable past
the clamp.
20. Verify that there is a small gap (approx. 1-3 mm) between
the Z-axis slider and the cap at the top of the Z-Drive.
Adjust the cable as necessary.
Section 3 The AutoSampler Module
Automation
3-9