4 User interface
4.3 Alarm screens
4.3.2 Alarm configuration screens
Weight alarm limits
The weight Critical Alarm limit is equal to the Max Weight interlock limit. The Critical
Alarm limit depends on the size of the vessel. If the Max Weight interlock limit is reached,
the pH regulation is deactivated. This alarm limit is displayed as read-only and cannot
be changed by the user.
See Section 7.3.7 Set alarm limits, on page 227 for instructions how to set the accessible
alarm limits. See Section 2.4 Xcellerex XDUO mixer interlocks, on page 43 for more infor-
mation about interlock limits.
Alarm functionality
When an alarm is issued, a warning icon
96
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
is displayed. The system continues to run.