GE Xcellerex XDM Series Operating Instructions Manual page 120

Table of Contents

Advertisement

5 Installation
5.2 Site requirements
Electrical power
Parameter
Supply voltage, mixer
Number of electrical outlets
Type of distribution system
Disconnect mixer
The mixer must be turned off and a tag-out device attached onto the I/O Cabinet during
installation. The system power switch must be set in the O OFF position and the power
cord must be disconnected.
Computer
An external computer is not required.
Temperature regulation
For temperature control of the jacketed XDM mixer or XDUO mixer, a temperature control
unit (TCU) is required. See the manufacturer´s manual for specified environmental condi-
tions and electrical power requirements for the TCU.
120
NOTICE
The temperature control unit must be disconnected before moving
the equipment.
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
Requirement
100-120 V AC, 50 or 60 Hz, 4.5 A
220-240 V AC, 50 or 60 Hz, 3.0 A
Two outlets for the following units:
I/O Cabinet
TCU
TN

Advertisement

Table of Contents
loading

Table of Contents