9 Troubleshooting
Problem
The agitator assembly does not
fit properly in place.
The agitator drive head rubs
against the agitator base plate
support and a grinding sound is
emitted. Plastic filings might be
generated.
Burst disc
Problem
Liquid is leaking from the vessel
jacket.
Local control panel
The information presented in this section is relevant for XDUO mixer only.
Problem
The local control panel does not
function.
The home screen does not show
up on the local control panel, al-
though the system is plugged in.
A button on local control panel
screen is not working.
A specific screen is not accessi-
ble from the local control panel.
266
Possible causes
The agitator is not properly
aligned.
Possible causes
Burst disc has ruptured.
Possible causes
Power cable is not connected.
The main power cable is dam-
aged.
The I/O Cabinet is broken.
The local control panel screen is
broken.
The mixer configuration is not
correctly set up.
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
Corrective action
Contact your GE representative
to schedule agitator alignment
check and maintenance.
Corrective action
Replace the burst disc. See Re-
place burst disc, on page 259 for
instructions.
Corrective action
Connect the power cable.
Contact your GE representative
for replacement cable.
Contact your GE representative
for possible replacement.
Contact your GE representative
for possible replacement.
Contact your GE representative
for assistance with configuration
adjustment.