6 Preparation
6.2 Stabilize the mixer vessel
6.2.1 Use jacking bolts
Step
3
4
5
134
Action
Turn down the top nut (1) with your fingers to the support plate, and tighten
it slightly with a wrench.
Turn down the securing nut (2) with your fingers to the top nut, and tighten
it slightly with a wrench.
Repeat steps 2 to 4 for all four jacking bolts installed on the mixer.
Result: The jacking bolts are engaged and the load cells are protected from
possible damage caused by movement.
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
2