9 Troubleshooting
Problem
Weight monitoring error:
Weight display on local con-
trol panel home screen dis-
plays a value outside of the
operational range
or
A weight alarm is triggered
when the mixer weight is
within the operational range.
272
Possible causes
Load cell has failed (physically
or electrically broken).
Physical obstruction of load cell
motion.
Jacking bolt is engaged.
Weight transmitter configuration
incorrect or corrupt.
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
Corrective action
Contact the metrology depart-
ment to perform a cornering test
and determine which load cell
has failed.
Verify that the mixer vessel legs
are free of any obstructions or
blocking objects.
Disengage the jacking bolt. See
Disengage jacking bolts, on
page 135 for instructions.
Note:
If the mixer was filled while the
jacking bolt was engaged, the
weight measurements of the
batch are not accurate.
Contact your GE representative.