GE Xcellerex XDM Series Operating Instructions Manual page 270

Table of Contents

Advertisement

9 Troubleshooting
Problem
Temperature control
unit (TCU) does not re-
ceive setpoint data:
the temperature is
not regulated
mixer vessel jacket
does not feel warm
TCU does not
sound as running
270
Possible causes
TCU is not powered.
There is no connection between the
I/O Cabinet and the PolyScience™
TCU.
There is no connection between the
I/O Cabinet and the Mokon™ TCU.
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
Corrective action
Switch on the power.
Check if the connection between the
I/O Cabinet and TCU is functioning
properly. Do the following:
1 Hold the clock button on the left
side of the TCU until HL is dis-
played (a few seconds).
2 Press the clock button repeatedly
until CC appears on the left dis-
play.
3 If the display message is yes, the
connection is working properly.
4 If the display message is no, the
connection is not functioning.
Check the connection cable be-
tween the TCU and PLC.
Check if the connection between the
I/O Cabinet and TCU is functioning
properly. Do the following:
1 Verify that the controller located
on the front of the TCU is pow-
ered.
2 Disconnect the cable between
the TCU and the I/O Cabinet.
3 Verify that the connector pins are
not bent or damaged. Check both
ends of the cable.
4 Inspect the whole length of the
cable for damage.
5 Verify that the controller is set up
to accept a remote input. See
Mokon manufacturer's manual
for this procedure.