If the adjustment of the temperature probe at the operating temperature does not bring
the system into tolerance, replace the temperature probe and repeat the calibration
procedure.
If the calibration of the new temperature probe does not bring the system into tolerance,
contact your GE representative for service.
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
Action
Type the calculated offsets into the Offset text boxes.
Tap BACK to return to the Transmitter Offsets screen.
Tap HOME to return to the home screen.
10 Reference information
10.8 Temperature probe calibration
293