Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
Action
Perform a visual inspection of the mated connection to confirm that the
sealing O-rings are in place and that they are not distorted.
NOTICE
If there is any indication that the connection might not
be aseptic, immediately clamp off this probe sheath
assembly on the short section of silicone tubing be-
tween the disposable bag and the female ACD connec-
tor.
6.4 Prepare and install the probes
6.4.4 Insert probes into the disposable bag
6 Preparation
175